Validation

Where the prototype stands and how it is being evaluated.

Early Bench Validation

Measured results

In bench testing against a reference gauge on a pelvis phantom, Rayfield's angle readout tracked true angular change to sub-degree precision — independently reviewed.

±0.1°
agreement with reference gauge (per 5° step)
0.05–0.09°
repeatability
R² ≈ 0.99
linearity — detects real change ~1:1
100%
tracking availability · 0 dropouts

Relative angular accuracy on a bench / phantom (inclinometer ground truth). Research prototype — absolute accuracy and pelvis registration are next steps. Not for clinical use.

Prototype Status

Rayfield is an early research prototype under active development.

Phantom Validation

Accuracy is assessed through phantom testing in controlled conditions.

Research Collaboration

We welcome research collaborations to evaluate spatial guidance approaches.

Clinical Advisory Network

Clinical engagement

Rayfield is receiving feedback from orthopedic clinicians and advisors in Canada, Israel and Ukraine as part of ongoing technical and clinical feasibility evaluation. The network supports workflow assessment, technical validation planning, and the development of future clinical evaluation protocols.

Technical Validation

Current activities

  • Orientation tracking validation
  • Phantom-based testing
  • Repeatability measurements
  • Workflow evaluation